The anti-snoring bed - a pilot study (2024)

  • Elisabeth Wilhelm1,
  • Francesco Crivelli1,2,
  • Nicolas Gerig1,3,
  • Malcolm Kohler4 &
  • Robert Riener1,5

Sleep Science and Practice volume4, Articlenumber:14 (2020) Cite this article

  • 12k Accesses

  • 2 Citations

  • 19 Altmetric

  • Metrics details

Abstract

Purpose

Avoiding supine position can reduce snoring in most habitual snorers. However, devices that restrict the sleeping position cause discomfort or disrupt sleep resulting in low compliance. Therefore, mechanisms, which lift the trunk of the user without disturbing sleep, have been proposed. We present the first study, which investigates whether individual interventions provided by beds with lifting mechanisms are able to stop snoring (success rate) and whether they reduce the snoring index (number of total snores divided by total time in bed) using a repeated measures design. In addition, we investigated whether the intervention is interfering with the subjective sleep quality.

Methods

Twenty-two subjects were observed for four nights (adaptation, baseline, and two intervention nights). During intervention nights, the bed lifted the trunk of the user in closed-loop manner. Subjects were divided in three groups (non-snorer, snorer one, and snorer two). Non-snorers were lifted by the bed at random time points during the night. In group snorer one, a stepwise increase of the bed inclination was compared with going directly to a randomly selected angle. In group snorer two, the influence of a small inclination angle (10 ) and a big inclination angle (20 ) was compared.

Results

Snoring was stopped successfully in 22% (small angle) and 67% (big angle) of the interventions. This did not lead to a significant reduction in the snoring index. The subjective sleep quality was not reduced by the intervention.

Conclusion

The anti-snoring bed is able to stop individual episodes of habitual snoring without reducing the subjective sleep quality.

Trial Registration

https://clinicaltrials.gov, no. NCT04053738, registered 12 August 2019 - Retrospectively registered, https://clinicaltrials.gov/ct2/show/NCT04053738.

Introduction

Habitual snoring is a widespread sleep problem (Deary et al. 2014), which does not only affect the health of the snorer (Endeshaw et al. 2013) but also the quality of life of the bed partner (Beninati et al. 1999). Intense snorers snore up to 245 times/hour (Cathcart et al. 2010). Snoring often occurs when the muscle tone drops while the snorer is lying in supine position. In this position, gravitational forces pull the soft tissue surrounding the upper airways in dorsal direction, thereby narrowing the airways. This causes audible flow turbulences in the upper airways (Gleadhill et al. 1991).

Therapeutic interventions force snorers to avoid supine position. The easiest way to implement positional therapy is to attach an obstacle, to the back of the snorer (Berger et al. 1997). More sophisticated devices include vibration (van Maanen and de Vries 2014) or sound (Cartwright et al. 1985) alarms raised when the snorer sleeps in supine position, or cushions, which push the head of the snorer sideward (Zhang et al. 2013). However, the discomfort needed to force the snorer to avoid supine position results in disrupted sleep and low user compliance (Bignold et al. 2009).

An alternative approach is to elevate the trunk of the user with cushions. McEvoy et al. reported that sleeping in seated position (trunk elevation of 60 ) has positive influences on the severity of obstructive sleep apnea (McEvoy et al. 1986). Recently it has been suggested to use adjustable bed frames to change the position of the sleeping snorer (Van Der Loos et al. 2003). Actuated beds such as the one used within this paper cost about 7000 Swiss francs and are therefore more expensive than conventional approaches for providing positional therapy. Commercially available versions such as the Partner Snore TM Technology (Sleep Number Corporation, Minneapolis, USA) elevate the trunk of the sleeping snorer if the bed partner indicates the presence of snoring via a push button (Sleep Number Corporation). Despite the fact that these beds are commercially available, little is known about the effect of such devices.

In this paper, we present the first systematic pilot study, which investigates the influence of automated trunk elevations on snorers. In absence of any evidence on the effect of different intervention parameters, we decided to conduct a pilot study, which investigates multiple options with three small groups. This pilot study aims on identifying the range of promising intervention options for future investigations. In particular, we were interested, which elevation angles would most likely reduce habitual snoring. Furthermore, we investigated how the bed influences the subjective sleep quality of the user.

Materials and methods

Setup

The mattress shape was adapted using a custom made intelligent anti-snoring bed (Fig.1), which is able to detect snoring sound and change the position of the user whenever snoring occurs (Crivelli et al. 2017; Wilhelm et al. 2017). The bed is triggered depending on the snoring index (SI) defined as number of snores detected within a given time frame. This time frame was set to one minute. The threshold that triggered bed motion was ten snores per minute for the first part of the study. After the interim analysis, this threshold was lowered to four snores per minute.

Measurement setup. a Postural interventions were provided using an automated bed, which can elevate the trunk of the sleeping user. A custom-made built-in microphone was used to trigger trunk elevation. In addition, a reference microphone and a IR-camera were used to monitor the user. b The inclination angles provided by the bed were 10 (P10), 15 (P15), and 20 (P20) provided at the location where the users hip joint is to be expected. At the beginning of the night on during episodes of silence the bed was set to neutral postion (P0)

Full size image

Subjects

Twenty-two participants (5 female, age 29.7 years SD: 16.0) participated in the study (Table1). Participants were not medication free. Alcohol and caffeine consumption was not restricted. Exclusion criteria were pregnancy, previously diagnosed sleep-related breathing disorders, chronic lower back pain, heart insufficiency that might impede sleeping in supine position, and inability to follow the procedures of the study. All subjects received a monetary compensation (CHF 25 per night) to cover travel costs. Daytime sleepiness was recorded using the Epworth Sleepiness Scale (ESS). Ten participants reached an ESS-score above 10, which indicates excessive daytime sleepiness (Johns 1991). Apnea hypopnea index (AHI) and oxygen desaturation index (ODI) were measured using a home-monitoring device (Apnea Link, purchased from ResMed). The scoring was done automatically by the ApneaLink software (purchased from ResMed). Two participants had an AHI and an ODI bigger than 5 during the adaptation night. One of these two participants also scored more than 10 points in the ESS. Therefore, we cannot exclude that these participants suffer from an undiagnosed sleep-related breathing disorder. In addition, participants were asked to report any breathing related co-morbidity. One participant reported being allergic to cats and having frequent breathing problems after smoking for about 6 years. Another participant reported being affected by allergic rhinitis (hay fever). Four participants reported experiencing symptoms of a cold on one or more days on which they participated in the study.

Full size table

Participants were divided into snorers and non-snorers based on the amount of snoring recorded with a home-monitoring device (Apnea Link, purchased from ResMed) during the first night. The threshold was set to 50 snores per night. This value was chosen with respect to the normal breath rate of a human subject (12 -20 breath/minute), someone who snores 50 times per night snores between 2.5 and 5 min in total. Interventions by the bed are only triggered after 1 minute of continuous snoring. Therefore, the chance to see interventions in snorers who snore less than 50 times per night would be very low. Furthermore, by excluding subjects with a snoring episode of less than 5 min we wanted to avoid overestimation of an overestimation of positive interventions. Positive interventions were defined as bad movements (which are triggered after 1 minute of snoring) that were able to stop snoring within 3 minutes time. Participants were recruited in two phases. The group snorer one was recruited between February and July 2017. The group snorer two was recruited from February to July 2018. All non-snorers recruited in both recruitment periods received the same interventions. Parameters for snorers were optimized after completion of the recruitment period.

One subject withdrew from the study. Four subjects did not receive the intervention in at least one night of the study due to technical problems. Three snorers did not trigger the bed. These datasets were excluded from the evaluation. Some subjects only triggered the bed in one of the experimental conditions. Nights in which the bed was not triggered were treated as missing data points.

Study protocol

All subjects spent four consecutive nights in the lab. Daytime activities were not restricted. Subjects went to bed at their habitual bedtime. The first night served as adaptation night. Night two to four were used as experimental nights. These nights were split into one baseline night and two nights with interventions. The order of the nights was randomized.

Within this study, we used four different elevation angles (0, 10, 15, and 20 degree referred to as P0, P10, P15, P20, respectively). The bed moved with an angular velocity of 1.5 /s. As depicted in Fig.2 the three groups got different interventions.

Interventions provided by the bed. Participants of the groups snorer one and non-snorer were provided with a stepwise intervention and a random angle intervention. During stepwise intervention, the bed was elevated upon the detection of snoring. If snoring stopped within 5 minutes, the bed went back to P0. Otherwise, the inclination was increased to the next step. During the random angle intervention, the bed went to one of the three inclination angles upon detection of snoring and waited in this position until snoring stopped. The group snorer two got two different interventions. During the small angle intervention, the bed moved to P10 upon detection of snoring and back to neutral when snoring stopped. During the big angle intervention, the bed went to P20

Full size image

Within the groups snorer one and non-snorer we wanted to investigate whether a stepwise increasing inclination angle (stepwise condition) is less disturbing than a sudden change of position (random angle condition) (Fig.2). Since Non-snorers would not trigger the bed with snoring sounds, interventions were provided to them at random time points throughout the night.

We compared the lowest angle (10 , P10) with the highest inclination angle (20 , P20) used within the group snorer two. Nights during which the bed only went to P10 are called small angle. The other nights are referred to as big angle.

Goal of the study

Within this study, we investigated, whether individual interventions of the bed are able to stop snoring episodes. In addition, we were interested, whether the overall amount of snoring would be reduced in intervention nights compared to baseline. Furthermore, we investigated whether the bed influences the subjective sleep quality.

Snoring activity was monitored using a portable home-monitoring device (Apnea Link plus, purchased from ResMed, Switzerland). Since audio based detection has been reported to be the most reliable way of snoring detection (Arnardottir et al. 2016), we also used a sound level meter (XL2 Audio and acoustic Analyzer, purchased from NTi Audio, Lichtenstein). Manual visual-audio scoring of the data collected with the noise level meter was performed using the XL2 data explorer.

The success rate of the individual intervention was derived by dividing the number of positive interventions through the total number of bed movements.

To investigate the effect on the total amount of snoring observed within one night, we calculated the SI using the lables of the software of the Apnea Link and the lables of the manual scoring. The SI was calculated by dividing the number of snores (S) through the recording time (trecord):

$$ {SI} =\frac{ {\sum\nolimits}_{i}S} {t_{record}}. $$

The sleeping position of the user was manually scored based on an infrared video. We distinguished between supine, prone, left, or right position and movement episodes.

Subjective Sleep Quality was assessed using the Groningen Sleep Quality Scale (GSQS) (Meijman et al. 1988). In addition, subjects were asked to indicate whether they think that an intervention occurred during the night, whether they woke up due to an intervention, and whether the intervention was disturbing their sleep. On the morning after the last experimental night, subjects were also asked to indicate which condition they preferred. The questionnaires were filled out within the first 10 minutes after wake up.

Statistical analysis

The data was analysed using Matlab r2018b (MathWorks, USA). Data was tested for normality using the Kolmogorow-Smirnow-Test. Since this test indicated a non-normal distribution, we used a non-parametric method. The Friedman-Test, which would be the non-parametric alternative for an ANOVA in repeated measures design, could not be used, since it does not accept missing values. Alternative tests, which do accept missing values, require a bigger sample size. Therefore, we used the Kruskal-Wallis Test to compare the values within each group. All results are reported as mean and range. The significance level was set to p=0.05.

We decided to look at each group individually during the analysis, because the number of bed movements differed in between groups.

Results

Number of interventions triggered by different groups

Non-snorers received 12 bed movements in both random angle and stepwise condition. The group Snorer one triggered the bed on average 1.5 (range: 0, 4) times and 2 (range: 0, 7) times in stepwise and random angle condition, respectively. In stepwise condition, all but one snorer stopped snoring when the bed reached P10. The other snorer stopped snoring when the bed was in P15. Averaged over all nights snorers of the group snorer one only encountered 6.3 (range: 0, 13) bed movements. This is only half of what the control group experienced. Therefore, we decided to investigate at each group individually.

Participants of the group snorer two triggered the bed on average 3 (range: 0, 11) and 2 (range: 0, 6) times in the small and big angle condition, respectively.

Effects of the postural intervention on snoring

In the group snorer one 60.0 % of the stepwise interventions and 57.0 % of the random angle interventions were successful. As depicted in Fig.3 the bed reduced the audio-based SI slightly in snorer one during the random angle condition (n.s.). In the data recorded with the home-monitoring device there was no reduction of snoring in intervention nights.

Effect of the automated bed on snoring. The average SI was once derived from manually scored audio-recording and once from the nasal cannula of the home-monitoring device. As depicted the bed was able to stop snoring with a success rate of 22.2 % to 66.6 % in different test conditions. However, the successful interventions did not lead to a reduction of the average SI which describes the total amount of snores divided by the time spent in bed

Full size image

In the group snorer two, small angle was provided 18 times with a success rate of 22.2 %. The intervention big angle was triggered 12 times with a success rate of 66.6 %. There was no statistical significant difference between the audio-based SI recorded during the different nights. The same applied for the SI measured with the home-monitoring device. However, both outcome measures showed a tendency towards a lower SI in the big angle condition (see Fig.3).

Effects of the postural intervention on subjective sleep quality

As depicted in Fig.4 the stepwise intervention seemed to decrease subjective sleep quality measured using the GSQS in the groups non-snorer and in snorer one (n.s.). Participants of both groups noticed in which nights interventions occurred. This result was significant in snorer one (\(p= 0.0498, \tilde {\chi }^{2}= 6.0000\), with a mean rank of 7.5, 3.0, and 3.0 for baseline, stepwise, and random angle respectively) but not in non-snorer. The questions whether subjects woke up due to an intervention and whether the bed did disturb the participants did not give significant results. Two out of four snorers and two out of six non-snorers did not indicate which condition they preferred. One out of six non-snorer chose the stepwise condition and one the random angle condition. All other participants indicated that they preferred the baseline condition.

Influence on subjective sleep quality. GSQS scores reach from 0 to 14 with 0 indicating good sleep and 14 indicating bad sleep. As depicted non-snorers slept worse during intervention nights. In group snorer one sleep quality improved slightly during the random angle condition. The group snorer two had a tendency to report better sleep quality in both intervention conditions. None of these results was significant

Full size image

In the group snorer two, we investigated the influence of small and big inclination angles. As depicted in Fig.4. there was no difference in GSQS. Subjects of snorer two noticed in which nights interventions occurred (\( p= 0.0481, \tilde {\chi }^{2}= 6.0701\)) with a mean rank of 13.67, 8.17, and 6.67 for baseline, small angle, and big angle, respectively. Furthermore, they reported that they had been awake due to the intervention (\( p= 0.0439, \tilde {\chi }^{2}= 6.2503\), mean rank of 5.3, 10.25, and 12.92 for baseline, small angle, and big angle, respectively). However, the question whether the intervention disturbed their sleep was not giving significant results. Three out of six participants of the group snorer two reported that the big angle condition was their favourite condition, two opted for baseline, and one for small angle condition.

Discussion

Within our study, snoring stopped in 60 %, 57 %, or 67 %, while a intervention in stepwise, random angle, or big angle, was provided, respectively. However, this did not lead to a significant reduction of the average SI. Eventually, the natural fluctuation of snoring covered the effect of the individual interventions. Within the same subject snoring frequency and the duration of snoring episodes can vary by 22 % and 33 %, respectively, in consecutive nights (Cathcart et al. 2010). Therefore, long-term studies might be needed to evaluate the effect of anti-snoring beds. While the natural variation of habitual snoring makes it hard to evaluate scientific studies, it also shows how important it is to provide interventions in closed-loop manner. The closed-loop approach allows the user to sleep in the sleeping position he prefers and will only intervene if snoring occurs. Therefore, it is much less restrictive than commonly used approaches.

In addition, our study had some limitations. In some nights, the nasal cannula was displaced due to movement. In these cases, the SI was only calculated for periods with valid measurements. Since the snoring activity varies throughout the different sleep stages this reduces the validity of this measure. The audio-based evaluation has been reported to be more reliable. However, due to the absence of a clear definition of snoring sound, this evaluation is influenced by the subjective experience of the expert scorer (Rohrmeier et al. 2014).

Furthermore, there are multiple causes for habitual snoring. We recruited snorers based on self-report. Therefore, we had a mixed population. So far, it has been suggested that anti-snoring beds can only be effective for position dependent snorers. One reason why we did not see significant changes of the SI could be that both our snorer groups did also contain snorers who were not position dependent.

Another limitation of our study is, that in non-snorers interventions were provided regardless of the sleeping position. Position dependent snorers would be expected to trigger the bed while lying in supine position. In the non-snorers in our study only 24% of the upward movements of the stepwise condition and 47% of the upward movements of the random angle condition occurred while the participants were lying in supine position. This might be due to the fact that only three out of the six participants of this groups spent more than 50% of the baseline night in supine position. By recruiting non-snorers who are mainly sleeping in supine position, one could investigate the effect of the intervention in this specific sleeping position.

In addition, the snoring detection algorithm of the anti-snoring bed did not recognize all types of snoring. Therefore, some snorers did not trigger the bed. Eight times the bed reacted to environmental noise. This implies that more robust snoring detection algorithms could decrease sleep disturbance caused by anti-snoring beds. Alternatively, the interventions provided by the bed could be triggered manually by an expert scorer. However, this solution would be very cost intensive and could therefore be only used in scientific studies and not as a long-term treatment for habitual snoring.

With respect to the effect of anti-snoring beds on subjective sleep quality, we saw a slight tendency towards an increased subjective sleep quality in snorers in conditions in which the bed moved to a target position and back. In the stepwise condition, the subjective sleep quality decreased. This might be because the bed moved more frequently in this condition. The effect of an intervention on the objective sleep quality can only be investigated using polysomnography. However, polysomnography is highly obtrusive and would reduce the subjective sleep quality of the participants. As reduced comfort is the most common reason for aborting conventional therapy approaches, our approach would only be a real alternative if the subjective sleep quality is at least similar to baseline condition. Therefore, we decided not to investigate objective sleep quality measures and did instead focus on the subjective sleep quality in our pilot study. In future studies polysomnography should be included to investigate whether the motion of the bed causes arousals or changes the sleep architecture.

Conclusion

In our Anti-snoring bed study, snoring stopped more frequently when interventions of the type big angle (20 ) were provided compared to interventions of the type small angle (10 ). This suggests that larger inclination angles might be more efficient. Furthermore, conditions with less frequent bed movements that cover a big range of motion seem to have less effect on subjective sleep quality than conditions with multiple smaller bed movements. The interventions provided by the bed did not lead to a significant reduction in subjective sleep. Further studies are needed to investigate whether Anti-snoring beds are a valuable alternative to conventional positional therapy.

Availability of data and materials

Video and audio data generated during the current study are not publicly available, because this data could not be anonymized. Tables containing the GSQS, the snoring indices, the manual scoring of the audio data of the snorers, the manual scoring of the video data of the snorers, and the scores obtained with the home-monitoring device are available on reasonable request in the ETH research collection repository under https://www.research-collection.ethz.ch/handle/20.500.11850/396977.

Abbreviations

BMI:

Body Mass Index

CHF:

Swiss Franc

ESS:

Epworth Sleepiness Scale

f:

Female

IR:

Infra red

m:

Male

n.s.:

not significant

P0:

Elevation angle 0

P10:

Elevation angle 10

P15:

Elevation angle 15

P20:

Elevation angle 20

S:

Number of snores

SD:

Standard deviation

SI:

Snoring index

t record :

Recording time

t wait :

Wait time

GSQS:

Groningen Sleep Quality Scale

AHI:

Apnea Hypopnea Index

ODI:

Oxygen Desaturation Index

References

  • Arnardottir, ES, Isleifsson B, Agustsson JS, Sigurdsson GA, Sigurgunnarsdottir MO, Sigurdarson GT, Saevarsson G, Sveinbjarnarson AT, Hoskuldsson S, Gislason T. How to measure snoring? a comparison of the microphone, cannula and piezoelectric sensor. J Sleep Res. 2016; 25(2):158–68.

    Article Google Scholar

  • Beninati, W, Harris CD, Herold DL, Shepard Jr JW. The Effect of Snoring and Obstructive Sleep Apnea on the Sleep Quality of Bed Partners. Mayo Clinic Proceedings. 1999; 74(10):955–8. https://doi.org/10.4065/74.10.955. http://www.sciencedirect.com/science/article/pii/S0025619611639918.

  • Berger, M, Oksenberg A, Silverberg D, Arons E, Radwan H, Iaina A. Avoiding the supine position during sleep lowers 24 h blood pressure in obstructive sleep apnea (osa) patients. J Hum Genet. 1997; 11(10):657–64.

    CAS Google Scholar

  • Bignold, JJ, Deans-Costi G, Goldsworthy MR, Robertson CA, McEvoy D, Catcheside PG, Mercer JD. Poor long-term patient compliance with the tennis ball technique for treating positional obstructive sleep apnea. J Clin Sleep Med. 2009; 5(05):428–30.

    Article Google Scholar

  • Cartwright, RD, Lloyd S, Lilie J, Kravitz H. Sleep position training as treatment for sleep apnea syndrome: a preliminary study. Sleep. 1985; 8(2):87–94.

    Article CAS Google Scholar

  • Cathcart, R, Hamilton D, Drinnan M, Gibson G, Wilson J. Night-to-night variation in snoring sound severity: one night studies are not reliable. Clin Otolaryngol. 2010; 35(3):198–203.

    Article CAS Google Scholar

  • Crivelli, F, Wilhelm E, van Sluijs R, Riener R. Actuated bed for a closed loop anti-snoring therapy. In: 2017 International Conference on Rehabilitation Robotics (ICORR). London: IEEE: 2017. p. 823–7.

    Google Scholar

  • Deary, V, Ellis JG, Wilson JA, Coulter C, Barclay NL. Simple snoring: not quite so simple after all?Sleep Med Rev. 2014; 18(6):453–62.

    Article Google Scholar

  • Endeshaw, Y, Rice TB, Schwartz AV, Stone KL, Manini TM, Satterfield S, Cummings S, Harris T, Pahor M, Study HA. Snoring, daytime sleepiness, and incident cardiovascular disease in the health, aging, and body composition study. Sleep. 2013; 36(11):1737–45.

    Article Google Scholar

  • Gleadhill, IC, Schwartz AR, Schubert N, Wise RA, Permutt S, Smith PL. Upper airway collapsibility in snorers and in patients with obstructive hypopnea and apnea. Am Rev Respir Dis. 1991; 143(6):1300–3.

    Article CAS Google Scholar

  • Johns, MW. A new method for measuring daytime sleepiness: the epworth sleepiness scale. Sleep. 1991; 14(6):540–5.

    Article CAS Google Scholar

  • McEvoy, RD, Sharp DJ, Thornton AT. The effects of posture on obstructive sleep apnea. Am Rev Respir Dis. 1986; 133(4):662–6.

    CAS PubMed Google Scholar

  • Meijman, T, de Vries-Griever A, De Vries G, Kampman R. The evaluation of the groningen sleep quality scale. Groningen: Heymans Bull (HB 88-13-EX). 1988; 2006.

  • Rohrmeier, C, Herzog M, Ettl T, Kuehnel TS. Distinguishing snoring sounds from breath sounds: a straightforward matter?Sleep Breathing. 2014; 18(1):169–76.

    Article Google Scholar

  • Sleep Number Corporation. FlexFit smart adjustable bases. https://www.sleepnumber.com/sn/en/c/bed-bases. Accessed 17 Feb 2019.

  • Van Der Loos, HM, Ullrich N, Kobayashi H. Development of sensate and robotic bed technologies for vital signs monitoring and sleep quality improvement. Auton Robot. 2003; 15(1):67–79.

    Article Google Scholar

  • van Maanen, JP, de Vries N. Long-term effectiveness and compliance of positional therapy with the sleep position trainer in the treatment of positional obstructive sleep apnea syndrome. Sleep. 2014; 37(7):1209–15.

    Article Google Scholar

  • Wilhelm, E, Crivelli F, Markic S, Kohler M, Pugliese F, Riener R. Ein intelligentes bett zur verminderung von schnarchen. Somnologie. 2017; 21(S2):52.

    Google Scholar

  • Zhang, J, Zhang Q, Wang Y, Qiu C. A real-time auto-adjustable smart pillow system for sleep apnea detection and treatment. In: 2013 ACM/IEEE International Conference on Information Processing in Sensor Networks (IPSN). Philadelphia: IEEE: 2013. p. 179–90.

    Google Scholar

Download references

Acknowledgements

We thank François Pugliese from Elite SA for his support throughout the project. In addition, we thank Michael Herold-Nadig and Marco Bader from ETH Zurich for their technical support, their advice and contribution during the design and the development of the devices. Furthermore, we thank Rachel van Sluijs, Antonino Crivello, Quincy Rondei, and Alexander Breuss for their support during the measurements. We also thank Jana Petr for her support during sound file scoring.

Funding

This work was supported by the Commission for Technology and Innovation (CTI) under grant No. 17988.1 PFIW-IW, the company Elite SA, and by the Swiss National Science Foundation (SNF) under grant No. 162809.

Author information

Authors and Affiliations

  1. Sensory-Motor Systems Lab, Department of Health Sciences and Technology, Eidgenossische Technische Hochschule Zurich, Zurich, Switzerland

    Elisabeth Wilhelm,Francesco Crivelli,Nicolas Gerig&Robert Riener

  2. CSM SA, Center Alpnach, Robotics & Automation, Alpnach Dorf, Switzerland

    Francesco Crivelli

  3. Bio-Inspired RObots for MEDicine-Lab (BIROMED-Lab), Department of Biomedical Engineering, University of Basel, Basel, Switzerland

    Nicolas Gerig

  4. Department of Pulmonology, University Hospital Zurich, Zurich, Switzerland

    Malcolm Kohler

  5. Spinal Cord Injury Center, University Hospital Balgrist, Zurich, Switzerland

    Robert Riener

Authors

  1. Elisabeth Wilhelm

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  2. Francesco Crivelli

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  3. Nicolas Gerig

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  4. Malcolm Kohler

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  5. Robert Riener

    View author publications

    You can also search for this author in PubMedGoogle Scholar

Contributions

The study was designed by FC, RR, and MK. Device and sensor setup were developed by FC, EW, NG, and RR. Data processing and analysis was done by NG, MK and EW. The manuscript was written by EW. The manuscript was critically reviewed by all co-authors. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Elisabeth Wilhelm.

Ethics declarations

Ethics approval and consent to participate

The experimental procedure was approved by the Ethics Committee of ETH Zurich, no. EK 2015-N-70. Written informed consent was obtained from all subjects.

Consent for publication

The person depicted in figure one signed consent stating that the picture can be used in this publication.

Competing interests

This study was conducted within an industry collaboration project with Elite SA. Elite SA contributed to the design of the device and the study, but not to data collection and analysis, decision to publish, or preparation of the manuscript.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

The anti-snoring bed - a pilot study (5)

Cite this article

Wilhelm, E., Crivelli, F., Gerig, N. et al. The anti-snoring bed - a pilot study. Sleep Science Practice 4, 14 (2020). https://doi.org/10.1186/s41606-020-00050-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41606-020-00050-2

Keywords

  • Habitual snoring
  • Intelligent bed
  • Positional therapy
The anti-snoring bed - a pilot study (2024)
Top Articles
Latest Posts
Recommended Articles
Article information

Author: Errol Quitzon

Last Updated:

Views: 6052

Rating: 4.9 / 5 (59 voted)

Reviews: 82% of readers found this page helpful

Author information

Name: Errol Quitzon

Birthday: 1993-04-02

Address: 70604 Haley Lane, Port Weldonside, TN 99233-0942

Phone: +9665282866296

Job: Product Retail Agent

Hobby: Computer programming, Horseback riding, Hooping, Dance, Ice skating, Backpacking, Rafting

Introduction: My name is Errol Quitzon, I am a fair, cute, fancy, clean, attractive, sparkling, kind person who loves writing and wants to share my knowledge and understanding with you.